To determine the intraexaminer repeatability and intersession reproduc translation - To determine the intraexaminer repeatability and intersession reproduc English how to say

To determine the intraexaminer repe

To determine the intraexaminer repeatability and intersession reproducibility of corneal deformation measurement using Scheimpflug noncontact tonometry (Corvis ST) on normal subjects.

Methods: Thirty-seven adults aged 20 to 48 years were invited to have their corneal deformation and curvature measurements taken using Corvis ST and Pentacam, respectively. Three consecutive measurements were taken for each instrument between 9:00 and 11:00 AM for intraexaminer repeatability analysis. Participants returned between 3:00 and 5:00 PM the same day for intersession reproducibility analysis.

Results: The most repeatable corneal parameter measured by Corvis ST was central corneal thickness ([CCT] ICC, 0.96; precision, 10.85 μm; repeatability, 15.34 μm; CV, 1.01%), followed by deformation amplitude ([DA] ICC, 0.80; precision, 0.08 mm; repeatability, 0.13 mm; CV, 4.33%), first applanation time ([1st A-time] ICC, 0.77; precision, 0.22 milliseconds; repeatability, 0.31 milliseconds; CV, 1.42), and intraocular pressure ([IOP] ICC, 0.75; precision, 1.39 mm Hg; repeatability, 1.97 mm Hg; CV, 4.98). Other parameters showed poor repeatability. The DA and 1st A-time showed good intersession reproducibility. The 95% limits of agreement were +0.13 to −0.13 mm for DA and +0.27 to −0.33 milliseconds for 1st A-time. The DA was negatively correlated with central corneal thickness (r = −0.53, p < 0.001) but not with corneal curvatures (flattest curvature, r = 0.13, p = 0.46; steepest curvature, r = 0.05, p = 0.75).

Conclusions: Corneal deformation parameters DA and 1st A-time were repeatable and reproducible. A thinner cornea was associated with a higher corneal deformation. Measurement of DA serves as an indicator of corneal biomechanical properties.
0/5000
From: -
To: -
Results (English) 1: [Copy]
Copied!
To determine the intraexaminer repeatability and intersession reproducibility of corneal deformation measurement using Scheimpflug noncontact tonometry (Corvis ST) on normal subjects.Methods: Thirty-seven adults aged 20 to 48 years were invited to have their corneal deformation and curvature measurements taken using Corvis ST and Pentacam, respectively. Three consecutive measurements were taken for each instrument between 9:00 and 11:00 AM for intraexaminer repeatability analysis. Participants returned between 3:00 and 5:00 PM the same day for intersession reproducibility analysis.Results: The most repeatable corneal parameter measured by Corvis ST was central corneal thickness ([CCT] ICC, 0.96; precision, 10.85 μm; repeatability, 15.34 μm; CV, 1.01%), followed by deformation amplitude ([DA] ICC, 0.80; precision, 0.08 mm; repeatability, 0.13 mm; CV, 4.33%), first applanation time ([1st A-time] ICC, 0.77; precision, 0.22 milliseconds; repeatability, 0.31 milliseconds; CV, 1.42), and intraocular pressure ([IOP] ICC, 0.75; precision, 1.39 mm Hg; repeatability, 1.97 mm Hg; CV, 4.98). Other parameters showed poor repeatability. The DA and 1st A-time showed good intersession reproducibility. The 95% limits of agreement were +0.13 to −0.13 mm for DA and +0.27 to −0.33 milliseconds for 1st A-time. The DA was negatively correlated with central corneal thickness (r = −0.53, p < 0.001) but not with corneal curvatures (flattest curvature, r = 0.13, p = 0.46; steepest curvature, r = 0.05, p = 0.75).Conclusions: Corneal deformation parameters DA and 1st A-time were repeatable and reproducible. A thinner cornea was associated with a higher corneal deformation. Measurement of DA serves as an indicator of corneal biomechanical properties.
Being translated, please wait..
Results (English) 3:[Copy]
Copied!
To determine the intraexaminer repeatability and intersession reproducibility of corneal deformation measurement using. Scheimpflug noncontact tonometry (Corvis ST) on normal subjects.

Methods: Thirty-seven adults aged 20 to 48 years were. Invited to have their corneal deformation and curvature measurements taken using Corvis ST, and Pentacam respectively.Three consecutive measurements were taken for each instrument between 9: 00 and 11: 00 AM for intraexaminer repeatability. Analysis. Participants returned between 3: 00 and 5: 00 PM the same day for intersession reproducibility analysis.

Results:? The most repeatable corneal parameter measured by Corvis ST was central corneal thickness ([], CCT ICC 0.96; precision 10.85, thermal m;? Repeatability 15.34, thermal m;CV, the 1.01%), followed by deformation amplitude ([], DA ICC 0.80; precision 0.08; repeatability 0.13, mm, mm; CV 4.33%),,, First applanation time ([], 1st A-time ICC 0.77; precision 0.22; repeatability 0.31, milliseconds, milliseconds; CV 1.42),,, And intraocular pressure ([], IOP ICC 0.75; precision 1.39 mm, Hg; repeatability 1.97 mm, Hg; CV 4.98). Other, parameters. Showed poor repeatability.The DA and 1st A-time showed good intersession reproducibility. The 95% limits of agreement were 0.13 to − 0.13 mm for. DA and 0.27 to − 0.33 milliseconds for 1st A-time. The DA was negatively correlated with central corneal thickness (R. Signed, 0.53 P < 0.001) but not with corneal curvatures (flattest curvature R =, =, 0.13 P 0.46; steepest curvature R = 0.05 P,,, = 0.75).

Conclusions:Corneal deformation parameters DA and 1st A-time were repeatable and reproducible. A thinner cornea was associated with. A higher corneal deformation. Measurement of DA serves as an indicator of corneal biomechanical properties.
Being translated, please wait..
 
Other languages
The translation tool support: Afrikaans, Albanian, Amharic, Arabic, Armenian, Azerbaijani, Basque, Belarusian, Bengali, Bosnian, Bulgarian, Catalan, Cebuano, Chichewa, Chinese, Chinese Traditional, Corsican, Croatian, Czech, Danish, Detect language, Dutch, English, Esperanto, Estonian, Filipino, Finnish, French, Frisian, Galician, Georgian, German, Greek, Gujarati, Haitian Creole, Hausa, Hawaiian, Hebrew, Hindi, Hmong, Hungarian, Icelandic, Igbo, Indonesian, Irish, Italian, Japanese, Javanese, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Korean, Kurdish (Kurmanji), Kyrgyz, Lao, Latin, Latvian, Lithuanian, Luxembourgish, Macedonian, Malagasy, Malay, Malayalam, Maltese, Maori, Marathi, Mongolian, Myanmar (Burmese), Nepali, Norwegian, Odia (Oriya), Pashto, Persian, Polish, Portuguese, Punjabi, Romanian, Russian, Samoan, Scots Gaelic, Serbian, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenian, Somali, Spanish, Sundanese, Swahili, Swedish, Tajik, Tamil, Tatar, Telugu, Thai, Turkish, Turkmen, Ukrainian, Urdu, Uyghur, Uzbek, Vietnamese, Welsh, Xhosa, Yiddish, Yoruba, Zulu, Language translation.

Copyright ©2025 I Love Translation. All reserved.

E-mail: